1. HOME
  2. Instruments&Products
  3. HPLC Columns
  4. Optical Isomer Separation Columns SUMICHIRAL
  5. SUMICHIRAL OA-8000 series

SUMICHIRAL OA-8000 series

Features

Usable for both normal phase systems and reversed phase systems
Able to separate a wide range of primary amines

Active site

Product specification

Product name SUMICHIRAL OA-8000
Chiral selectors Pseudo-18-crown-6-ether
Support High purity silica gel
Particle size 5μm
Supporting mode Chemically bonded
Recommended mobile phase [Normal phase] Hexane/ethanol/trifluoroacetic acid
[Reversed phase] perchloric acis aqueous solution/acetonitrile
Target compounds Primary amines

Product list

Product name Particle size
(μm)
Internal diameter(mm) Length
(mm)
Product code
SUMICHIRAL OA-8000 5 2.0 150 C-8001-2015W
250 C-8001-2025W
4.0 10 C-8001-4001W
4.6 150 C-8001-4615W
250 C-8001-4625W
  • Note: Those shown in red are standard size for analysis.
    Internal diameter of 4 mm and length of 10 mm is a guard column.
    SUMIPAX Filter can be used in place of a guard column.

Applications

Amino acids/amino acid derivatives

SUMICHIRAL OA-8000(5μm, 4.6mm i.d.×250mm)
hexane / ethanol / trifluoroacetic acid / water(70 : 30 : 0.5 : 0.5)
UV 254 nm | 0.7 mL / min

Amino acids/amino acid derivatives

SUMICHIRAL OA-8000(5μm, 4.6mm i.d.×250mm)
hexane / ethanol / trifluoroacetic acid(70 : 30 : 0.5)
UV 223 nm | 0.7 mL / min | 10μL (12.5 mg / mL)

Amino acids/amino acid derivatives

SUMICHIRAL OA-8000(5μm, 4.6mm i.d.×250mm)
hexane / ethanol / trifluoroacetic acid / water(70 : 30 : 0.5 : 0.5)
UV 254 nm | 0.7 mL / min | 2μL (5 mg / mL)

Amino acids/amino acid derivatives

SUMICHIRAL OA-8000(5μm, 4.6mm i.d.×250mm)
perchloric acid (pH 2.0) / acetonitrile (80 : 20)
UV 254 nm | 0.7 mL / min | 1μL (5 mg / mL)

Primary amines, amino alcohols

SUMICHIRAL OA-8000(5μm, 4.6mm i.d.×250mm)
hexane / ethanol / trifluoroacetic acid (70 : 30 : 0.5)
UV 254 nm | 0.7 mL / min

Primary amines, amino alcohols

SUMICHIRAL OA-8000(5μm, 4.6mm i.d.×250mm)
perchloric acid (pH 2.0) / acetonitrile (70 : 30)
UV 254 nm | 0.8 mL / min | 1μL (1 mg / mL)

Contact Us for HPLC Columns

Please contact us for any information on related to HPLC columns.

Materials Division (Column Sales Team)